基于ACR TI⁃RADS 分析S⁃Thyroid 和超声医师评估甲状腺结节的一致性

周跃1 李卫民1 范晓芳1 高启2 吴文娟1 贾磊1    

  1. 江南大学附属医院超声医学科(江苏无锡 214000);2 东南大学附属中大医院超声医学科(南京 210009)

  • 出版日期:2022-09-25 发布日期:2022-09-25
  • 通讯作者: 李卫民 E⁃mail:1005342597@qq.com
  • 基金资助:
    无锡市妇幼健康推广项目(编号:FYTG202103)

Consistency between S⁃Thyroid and sonographer in evaluating thyroid nodules with ACR TI⁃RADS

ZHOU Yue*,LI Weimin,FAN Xiaofang,GAO Qi,WU Wenjuan,JIA Lei.   

  1. Department of UltrasoundAffiliated Hospital of Jiangnan UniversityWuxi 214000China 
  • Online:2022-09-25 Published:2022-09-25
  • Contact: LI Weimin E⁃mail:1005342597@qq.com
  • Supported by:


摘要:

目的 根据美国放射学会制定的甲状腺影像报告与数据系统(ACR TI⁃RADS),分析甲状腺 自动扫查助手(S⁃Thyroid)和超声医师对甲状腺结节超声特征评估的一致性和两者的诊断效能。方法 取甲状腺结节患者 253 例共 276 个结节,分别由甲状腺扫查助手(S⁃Thyroid)和超声医师对甲状腺结节的超 声特征进行评估,采用 Cohen′s Kappa 检验分析 S⁃Thyroid 和超声医师对甲状腺结节超声特征评估的一致 性,并以手术病理为金标准,绘制 ROC 曲线,分析二者的诊断效能。结果 276 个结节中,良性 89 个,恶性 187 个。S⁃Thyroid 和超声医师对甲状腺结节形态、内部结构以及回声的评估一致性强(Kappa 值分别为 0.973、0.886、0.805),对强回声灶和边缘评估的一致性中等(Kappa 值分别为 0.613、0.597)。在诊断效能方面,S⁃Thyroid和超声医师对甲状腺恶性结节的敏感度、特异度分别为82.01% vs. 86.24%,86.21% vs. 83.91% 差异均无统计学意义(均 > 0.05);曲线下面积(AUC)分别为 0.835(95%CI:0.801~0.926)、0.891(95%CI 0.846 ~ 0.936),差异无统计学意义(= 0.913,> 0.05)。结论 根据ACR 评分标准,S⁃Thyroid 和超声医师 对甲状腺结节超声特征的评估有着较好的一致性,且具有较高的诊断效能,值得进一步推广和应用。


关键词:

甲状腺影像报告与数据系统, 甲状腺自动扫查助手, 超声, 甲状腺结节, 一致性

Abstract:

Objective To investigate the consistency of S⁃Thyroid and sonographer in the evaluating char⁃ acteristics of thyroid nodules according to American College of Radiology Thyroid Imaging Report and Data System (ACR TI⁃RADS)and compare their diagnostic efficiency. Methods A total of 253 patients with 276 thyroid nod⁃ ules were enrolled in the study. The ultrasonic characteristics of thyroid nodules were evaluated by S⁃Thyroid and sonographers respectively. Cohen's kappa test was used to evaluate the consistency between S⁃Thyroid and sonogra⁃ phers in the diagnosis. Taking surgical pathology as gold standard,ROC curves were drawn and diagnostic efficacy was analyzed. Results In terms of ultrasonic characteristics,there were high consistency in morphology and inter⁃ nal structure(Kappa values were 0.973,0.886,0.805,respectively)and medium consistency in hyperechoic foci and edges(Kappa values were 0.613 and 0.597,respectively). Meanwhile,the sensitivity and specificity of S⁃Thy⁃ roid and sonographers were 82.01% vs. 86.24% and 86.21% vs. 83.91%,respectively. There were no significant dif⁃ ference(P > 0.05). The area under the curve was 0.835(95%CI:0.801 ~ 0.926)and 0.891(95%CI:0.846 ~ 0.936),respectively,and there was no significant difference between them(Z = 0.913,P > 0.05). Conclusion According to the criteria of ACR TI⁃RADS,there are good consistency and high diagnostic efficiency in the evalua⁃ tion of thyroid nodules between S⁃Thyroid and sonographers in evaluating thyroid nodules. Therefore,S⁃Thyroid is worthy of further popularization and application in clinic.

Key words:

ACR TI?RADS, S?Thyroid, ultrasound, thyroid, consistency