 Volumetric measurement of post ⁃stroke hemiplegic edematous hand by 3⁃D scanning and water displace⁃ ment

LIN Hua,WANG Nian,LIU Zhilan,LI Guo,YUAN Jing.   

  1. Department of Rehabilitation MedicineShang⁃ hai Fourth Rehabilitation HospitalShanghai 200040China
  • Online:2021-09-10 Published:2021-09-10
  • Contact: YUAN Jing E⁃mail:13524530167@163.com
  • Supported by:


摘要:

目的 评价三维扫描法和传统水置换法应用于脑卒中患者双手体积的测量方法和结果。 方法 选择2019年9月至2021年3月在上海市第四康复医院住院的100例首次发病的脑卒中偏瘫患手水肿患者,采用手持非接触式三维扫描仪对其进行双手扫描,获取手部三维点云模型,通过计算机软件对双手 模型进行去噪等处理,以桡、尺骨茎突连线为基线测量左右手体积;同时采用传统水置换法测量这 100 脑卒中偏瘫患手水肿患者的双手体积。对两种测量方法得到的测量结果进行统计学分析。结果 采用 两种方法测量100例脑卒中偏瘫患手水肿患者得到的健侧和患侧手体积结果无明显差异(健手:t = 1.378 P = 0.171;患手:t = 1.973,P = 0.051);三维扫描法测得的体积与水置换法测得的体积存在明显相关性(健 手:r = 0.992,P < 0.001;患手:r = 0.997,P < 0.001);Bland⁃Altman 图示两种测量方法测得的体积数据在 95%的一致性范围内,一致性良好。结论 非接触式三维扫描法测量脑卒中偏瘫患手体积与传统水置换 法测量手体积结果一致,三维扫描获得的手部数据模型可以方便地进行分部测量和分析,可为研究患手 水肿恢复状况提供强有力的研究工具。

关键词: 三维扫描法,  , 水置换法,  , 脑卒中偏瘫患手水肿,  , 手体积

Abstract:

Objective To analyze the volumetric measurement of post⁃stroke hemiplegic edematous hands using 3⁃D scanning and water displacement. Methods Hand volumes of 100 post⁃stroke hemiplegic patients were measured using a non⁃invasive 3⁃D scanner. We defined the hand mark as the line between the ulnar styloid and the radius styloid,then 3⁃D forms of the hands were captured by the 3⁃D scanning system. Traditional water displacement was used for volumetric measurement as well. Statistical analysis was applied to compare the hand volumetric results between the two measurement methods. Results There was no significant difference between two methods of hand volume measurement(asymptomatic hand:t = 1.378,P = 0.171;edematous hand:t = 1.973 P = 0.051). A significantly strong positive correlation was observed between the asymptomatic hand(r = 0.992 P < 0.001)and edematous hand volume(r = 0.997,P < 0.001)calculated via 3⁃D scanning and water displacement. The Bland⁃Altman analysis showed that the two measuring volume methods were within the 95% consistency limits of agreement(LOA). Conclusion The contactless three⁃dimension scanning method has the same effect in mea⁃ suring post⁃stroke edematous hand volumes when compared with the traditional method of water displacement. It can be used toconduct an accurate volumetric analysis of an edematous hand and provide reference for future research on the recovery of hand edema.